Digitron 2082P digital logging pressure meter
Digitron 2082P digital logging pressure meter
2 Bar (4 Bar) datalogging
- Operating range : 0-2 Bar
- Over range : 4 Bar
- Accuracy : 0,15% rdg
- Datalogging
EUR 394.00
excl. VAT
excl. VAT
Article code: 170-2082
- Product info
Easy to use, reliable and built to last, the 2082P digital manometer has been specifically developed in order to meet the performance and cost demands of today's industrial and HVAC professionals.
A range of LCD digital manometers for users who want simple and accurate pressure measurement. The manometers are suitable for air or liquid pressure measurement and find applications in heating and ventilation, gas flow, hydraulic systems, leak detection, process plant pressure checking and pump performance tests. Available in many different range options.
- Display : large LCD with backlight
- Operating range : 0-2 Bar
- Over range : 4 Bar
- Accuracy : 0,15% rdg + 0,15% fs + 1 digit
- Scales : mbar, bar, PSI, inH2O, inHg, mmH2O, mH2O, mmHg, cmHg, mHg, Pa & kPa
- Operation temperature : -10°C to +50°C
- Battery type : 2 x 1,5V type AA
- Battery life : 200 hours
- Datalogging
- Low battery check : symbol appears on display
- Standard functions : backlight, verlichting display, autoranging, Hold
- Overrange / Underrange : "out shows on display
- Environmental Specifications : IP67 (as std) with hose connected
- Auto switch-off time : 12 minutes
- Automatic zeroing
- Size : 155 x 67 x 40 mm
- Weight : 180 g
Pressure Scales
mbar | 0,0 ... 199,9 mbar + 0 ...19999 mbar |
Pa | 0,00 ... 19,99 kPa + 0,0 ... 19,9 kPa |
in H2O | 0,00 ... 199,9" + 0 ... 804" |
m H2O | 0,00 ... 20,43 m |
in Hg | 0,00 ... 59,00" |
m Hg | 0,00 ... 19,99 cm + 0,0 ... 150,0 cm |
PSI | 0,00 ... 29,00 psi |
Functions 2082P :
- backlight
- operating range
- over range
- out of range
- zeroing
- hold
- smoothing
- units of pressure
- range lock
- Max/Min
- averaging
- leak testing
- datalogging
Functions 2xxxP
Logging Mode, for storage, retrieval and output of readings, autoranging displays for precise readings, min/max, everage, logging and leak testing.
Datalogging
Logging functions preset Interval Logging lets you determine the frequency at which reading are taken. One the instrument has been set, readings are taken automatically at the interval you decide (from once every minute, to once every 24 houts), providing an excellent means of trouble-shooting and examining trends over a given period of time.
With Preset Interval Logging, you really can "set it, forget it", as the instrument will continue logging until a maximum of 250 readings are taken or your data is downloaded.
Data can be vieuwd and analysed direct from the magnified display. Data can be downloaded to a PC or Epson compatible printer via Digitron DigiLink and can be analysed using DigiLow Windows software.
Accesories
- 2000BT protectiv rubber boot
- 6 mm Nylon Tubing
- ... meter Nylon Tubing
- 6 mm Nylon "T" piece
- carry case
- Calibration Certificate
- Digilink logging interface
- software
- pressure pump max 5 bar
- vacuum pump max -750 mbar